YSI NiCaVis 701 IQ NI User Manual

Page 31

Advertising
background image

NiCaVis 70x IQ NI

Measuring / Operation

4 - 7

ba76132e01

03/2013

Default values are marked in bold.

Calibration evaluation

The entered calibration data are checked for plausibility by the system.

The check can have the following results:

Information on the contents and structure of the log book and how you

can call it up is given in the L

OG

BOOK

chapter of the IQ S

ENSOR

N

ET

system operating manual.

The user calibration is not recorded in the calibration history. You can,

however, view the current value pairs in the

Settings of sensors and

diff. sensors

menu at any time.

Menu item

Settings

Explanations

Cal - # raw value 1

0.00 ... 5000.00

Raw value of the first value pair

(lower concentration).

For single-point calibration, enter 0.

Cal - ref. value 1

0.00 ... 5000.00

Reference value of the first value pair

(lower concentration).

For single-point calibration, enter 0.

Cal - # raw value 2

0.00 ... 10.00 ... 5000.00 Raw value of the second value pair

(higher concentration) or raw value of the

single-point calibration.

Cal - ref. value 2

0.00 ... 10.00 ... 5000.00 Reference value of the second value pair

(higher concentration) or reference value

of the single-point calibration.

Display

Meaning/actions

Measured value dis-

play

Successful new valid calibration value pairs

were entered for the sensor.

"----"

Calibration value pairs not accepted. Sen-

sor is blocked for measurement. A log book

message is generated.

Causes:

Ascending order not observed.

At least one value is outside the measur-

ing range.