3 connecting the sensor to the iq sensor net, Connecting the sensor to the iq s – YSI NiCaVis 701 IQ NI User Manual

Page 15

Advertising
background image

NiCaVis 70x IQ NI

Commissioning

3 - 5

ba76132e01

03/2013

3.3.3 Connecting the sensor to the IQ S

ENSOR

N

ET

Connection cable

A sensor connection cable of the SACIQ or SACIQ SW type is required

to connect the sensor. The cable is available in different lengths.

Compared to the standard model SACIQ, the SACIQ SW sensor

connection cable is optimized regarding its corrosion resistance in

seawater and brackish water and adapted for use in conjunction with

the NiCaVis 70x IQ NI . Information on this and other IQ S

ENSOR

N

ET

accessories is given in the YSI catalog and on the Internet.
How to connect the SACIQ (SW) sensor connection cable to the termi-

nal strip of an MIQ module is described in chapter 3 I

NSTALLATION

of the

IQ S

ENSOR

N

ET

system operating manual.

Are the plug connections

dry?

Before connecting the sensor and sensor connection cable, please

make sure the plug connections are dry. If moisture gets into the plug

connections, first dry the plug connections (dab them dry or blow them

dry using compressed air).
Note
Do not suspend the sensor on the sensor connection cable but use the
suitable mounting equipment. Information on this and other
IQ S

ENSOR

N

ET

accessories is given in the YSI catalog and on the In-

ternet.