NiCaVis 70x IQ NI

Measuring / Operation

4 - 11

ba76132e01

03/2013

Fig. 4-5 Putting the sensor sleeve on the sensor

Carry out the intended sensor check or

Zero adjustment

directly after

filling the sensor sleeve. If you wait longer, the result of the Sensor

check/

Zero adjustment

will be affected.

Carrying out the sensor

check or

Zero adjustment

9.

Put the sensor in a horizontal position on a firm and vibration-

free surface. Turn the filling opening of the sensor sleeve

upward (Fig. 4-5).

10.

Rinse the measuring chamber with ultrapure water at least 3

times.

11.

For

Sensor check (H2O dest.)

or

Zero adjustment

:

Fill the measuring chamber with ultrapure water.

or

For

Sensor check (standard)

:

Fill the measuring chamber with standard solution (see sec-

tion 6).

12.

If there are any air bubbles on the measuring windows remove

them by gently knocking on the sensor sleeve with a bar.

Measuring
chamber

Sensor
sleeve

13.

Select the required option with <> and <OK>:

Sensor check (H2O dest.)

Zero adjustment

Sensor check (standard)

14.

Confirm with <OK>.

A multipart checklist appears. Use it to check once more

whether all preparations have been completed.