NiCaVis 70x IQ NI

Commissioning

3 - 11

ba76132e01

03/2013

Response time t90

10 ... 60 min

Response time of the signal smoothing.

Depending on the sample matrix, mea-

sured values can vary more or less

strongly (e.g. due to foreign bodies or

air bubbles).

The signal filter (here, the

Response

time t90

)

reduces the limits of variation

of the measured values.

The

Response time t90

specifies the

time after which 90 % of a signal

change are displayed.

This menu item is only available with:

Signal smoothing

=

On

The specified time is used as a guide

value for low (10 min) to high (60 min)

signal smoothing and can deviate from

the value that was set.

Cleaning mode

On

Off

Switching on/off the automatic, time-

controlled cleaning function.

Cleaning type

Ultrasound

Air

Ultrasound+air

Cleaning methods

This menu item is only available with:

Cleaning mode

=

On

The duration of the cleaning procedure

is defined in the sensor for all cleaning

types.

Note: The duration of the cleaning pro-

cedure with air can be adjusted for the

linked relay in the menu,

150 - Settings

of outputs and links

/

Relay function

/

Sensorgesteuert

/

Pulsdauer

.

Please note the software requirements

for the MIQ/CR3, MIQ/R6, MIQ/CHVP-

lus, DIQ/S 182 relays (see section 3.1).

Save and quit

-

The system confirms the saving of the

settings and the display switches to the

next higher level.

Quit

-

The display switches to the next higher

level without saving the new settings.

Menu item

Settings

Explanations