YSI NiCaVis 701 IQ NI User Manual

Page 34

Advertising
background image

Measuring / Operation

NiCaVis 70x IQ NI

4 - 10

ba76132e01

03/2013

Preparing the sensor

check or

Zero adjustment

Prepare the sensor check or

Zero adjustment

as follows:

Fig. 4-4 Maintenance condition

By activating the maintenance condition, linked outputs remain in their

present state. The cleaning and the determination of measured values

are switched off. The measured value display shows a flashing, invalid

measured value ("----").

1.

Switch to the measured value display with <M>.

2.

Use <> to select the UV sensor.

3.

Call up calibration with <C>. The

Maintenance condition:

Linked outputs are frozen.

window appears.

4.

Confirm the maintenance condition with <OK>.

5.

Take the sensor out of the test sample.

6.

Remove the blind plug or unscrew the compressed air connec-

tion from the sensor.

7.

Thoroughly clean the sensor enclosure, sensor sleeve and the

compressed air channels (see section 5.1 M

AINTENANCE

).

To finish the cleaning procedure, thoroughly rinse the entire

sensor with ultrapure water.

8.

Mount the cleaned sensor sleeve:

Loosen both coupling rings on the sensor sleeve.

Put the sensor sleeve on the sensor and align it:
– The measuring gap is in the center of the sensor sleeve
– The filling opening of the sensor sleeve and the measur-

ing gap opening of the sensor are offset by 90°.

Tighten both coupling rings on the sensor sleeve by hand.