Measuring / Operation

NiCaVis 70x IQ NI

4 - 2

ba76132e01

03/2013

4.2

Measuring operation

CAUTION

Never look into the measuring window or put objects into the measur-

ing window during operation! The emitted UV radiation can damage the

eyes!

In the case of authorized use, inadvertent eye contact with the light

beam is not possible.

Measured value display

Fig. 4-2 NiCaVis 70x IQ NI

in the measured value display with main sensor and vir

-

tual sensor

Factors affecting the

measured value

The following factors can adversely affect the measurement quality:

The measuring gap is not sufficiently filled and flowed through by the

test sample

Foreign bodies and air bubbles are in the measuring gap

Extraordinary changes (interferences) of the sample matrix

Contaminated measurement windows due to difficult waste water

matrix

Damaged measuring windows

Lacking quality of the zero adjustment, e. g. due to contaminated

ultrapure water

The user calibration is not representative (or erroneous)

Raw value,

identifiable by #

Measured value