10 appendix: glossary, 10 appendix: glossary -1 – YSI NiCaVis 701 IQ NI User Manual

Page 61

Advertising
background image

NiCaVis 70x IQ NI

Appendix: Glossary

10 - 1

ba76132e01

03/2013

10 Appendix: Glossary

Resolution

Smallest difference between two measured values that can be

displayed by a meter.

BSB, BOD

Abbreviation for the "Biochemical Oxygen Demand". The biochemical

oxygen demand of the amount of oxygen required by microorganisms

in order to aerobically decompose organic substances in an aqueous

solution.

CSB, COD

Abbreviation for the "Chemical Oxygen Demand". The chemical oxy-

gen demand is the equivalent amount of oxygen that is required for

the complete oxidation of the organic (major part) and anorganic

(minor part) substances in an aqueous solution.

DOC

Abbreviation for dissolved organic carbon. Concentration of

dissolved, organically bound carbon in water.

Absorbance

(Absorption)

Attenuation of the light beam when it penetrates the sample.

Adjusting

To manipulate a measuring system so that the relevant value (e. g. the

displayed value) differs as little as possible from the correct value or

a value that is regarded as correct, or that the difference remains

within the tolerance.

Calibration

Comparing the value from a measuring system (e. g. the displayed

value) to the correct value or a value that is regarded as correct.

Often, this expression is also used when the measuring system is

adjusted at the same time (see adjusting).

Calibration value pair

Value pair consisting of the raw value measured by the UV sensor,

and a reference value measured e. g. by a laboratory determination.

The calibration value pairs are the result of the user calibration.

Concentration

Mass or amount of a dissolved substance per volume, e. g. in g/L or

mol/L.

Measuring system

The measuring system comprises all the devices used for measuring,

e. g. measuring instrument and sensor. In addition, there is the cable

and possibly an amplifier, terminal strip and armature.

Measured parameter

The measured parameter is the physical dimension determined by

measuring, e. g. pH, conductivity or D.O. concentration.

Measured parameter

(correlated)

With the measuring method used, a correlated measured parameter

is not determined directly. A correlated measured parameter can,

however, be derived from a determinable measured parameter. The

precondition for this is that there is a reproducible interrelationship

(correlation) between the determinable and correlated measured

parameter.

Test sample

Designation of the test sample ready to be measured.

Advertising
This manual is related to the following products: