2 measuring the temperature, Measuring the temperature – Xylem MultiLab 4010-2 User Manual

Page 44

Advertising
background image

ORP voltage

MultiLab 4010-2

44

ba76142e01

05/2013

Criteria for a stable

measured value

The Stability control function checks whether the measured values are stable
within the monitored time interval.

The minimum duration until a measured value is assessed as stable is the
monitored time interval. The actual duration is mostly longer.

6.1.2 Measuring the temperature

For reproducible ORP measurements, it is essential to measure the tempera-
ture of the test sample.
When operating a sensor without integrated temperature sensor, you first have

1.

Freeze the measured value with <AR>.
The [HOLD] status indicator is displayed.
The HOLD function is active.

You can terminate the Stability control function and the HOLD func-
tion with <AR> or <M> at any time.

2.

Using <ENTER>, activate the Stability control function manually.
The [AR] status indicator appears while the measured value is assessed
as not stable. A progress bar is displayed and the display of the mea-
sured parameter flashes.
As soon as a measured value meets the stability criteria, it is frozen. The
[HOLD][AR] status indicator is displayed, the progress bar disappears
and the display of the measured parameter stops flashing.
The current measurement data is output to the interface. Measurement
data meeting the stability control criterion is marked by AR.

You can prematurely terminate the Stability control function manu-
ally with <ENTER> at any time. When the Stability control function
is prematurely terminated, the current measurement data are out-
put to the interface (PC, USB memory device or USB printer) with-
out AutoRead info.

3.

Using <ENTER>, start a further measurement with stability control.
or
Release the frozen measured value again with <AR> or <M>.
The [AR] status display disappears. The display switches back to the pre-
vious indication.

Measured
parameter

Time interval

Stability in the time
interval

ORP

15 seconds

Δ

: Better than 0.3 mV

Temperature

15 seconds

Δ

: Better than 0.5 °C