2 measuring the temperature, 2 calibration, 1 why calibrate – Xylem MultiLab 4010-2 User Manual

Page 48: 2 when to calibrate, 3 calibration procedures, Measuring the temperature, Calibration, Why calibrate, When to calibrate, Calibration procedures

Advertising
background image

Dissolved oxygen (D.O.)

MultiLab 4010-2

48

ba76142e01

05/2013

Criteria for a stable

measured value

The Stability control function checks whether the measured values are stable
within the monitored time interval.

The minimum duration until a measured value is assessed as stable is the mon-
itored time interval. The actual duration is mostly longer.

7.1.2 Measuring the temperature

For reproducible D.O. measurements, it is essential to measure the tempera-
ture of the test sample.
IDS D.O. sensors measure the temperature with a temperature sensor inte-
grated in the IDS sensor.

7.2

Calibration

7.2.1 Why calibrate?

D.O. sensors age. This changes the zero point (asymmetry) and slope of the
D.O. sensor. As a result, an inexact measured value is displayed. Calibration
determines and stores the current values of the zero point and slope.

7.2.2 When to calibrate?

When the calibration interval has expired

When your accuracy requirements are especially high

Routinely within the framework of the company quality assurance

7.2.3 Calibration procedures

The MultiLab 4010-2 provides 2 calibration procedures:

3.

Using <ENTER>, start a further measurement with stability control.
or
Release the frozen measured value again with <AR> or <M>.
The [AR] status display disappears. The display switches back to the
previous indication.

Measured parameter

Time interval

Stability in the time
interval

D.O. concentration

20 seconds

Δ

: better than 0.03 mg/l

D.O. saturation

20 seconds

Δ

: better than 0.4 %

D.O. partial pressure

20 seconds

Δ

: Better than 0.8 mbar

Temperature

15 seconds

Δ

: Better than 0.5 °C

Advertising