2 orp measurement settings, Orp measurement settings – Xylem MultiLab 4010-2 User Manual

Page 65

Advertising
background image

MultiLab 4010-2

Settings

ba76142e01

05/2013

65

Setting the calibra-

tion interval

The calibration interval is set to 7 days in the factory.
You can change the interval (1 ... 999 days):

9.2

ORP measurement settings

Settings

The settings are made in the menu for measuring settings of the ORP measure-
ment. To open the settings, display the required measured parameter in the
measured value display and press the <ENTER> key. After completing the set-
tings, switch to the measured value display with <M>.
Default settings are printed in bold.

1.

Open the menu for measurement settings with <ENTER>.

2.

In the Calibration / Calibration interval menu, set the calibration interval
with <><>.

3.

Confirm the setting with <ENTER>.

4.

Quit the menu with <M>.

Menu item

Possible setting

Explanation

mV resolution

0.1
1

Resolution of the mV display

Reset

-

Resets all sensor settings to the
delivery condition (see section
9.6.1 R

ESETTING

THE

MEASUREMENT

SETTINGS

, page 73).