13.3 Conductivity
Error message,
OFL
Error message,
Error
– Measured temperature value outside
the operating conditions
(display of OFL/UFL instead of a
temperature value)
– Keep to the temperature range
for the test sample
– Defective sensor
– Calibrate
– Change sensor cap
– Replace sensor
More information and instructions on cleaning and exchange of
sensors are given in the documentation of your sensor.
Cause
Remedy
Cause
Remedy
– Measured value outside the measur-
ing range
– Use suitable IDS conductivity
sensor
Cause
Remedy
– Sensor contaminated
– Clean the sensor and replace it
if necessary
– Calibration solution not suitable
– Check the calibration solutions