6 reset, 1 resetting the measurement settings, Reset – Xylem MultiLab 4010-2 User Manual
Page 73: Resetting the measurement settings, Section 9.6.1 r, Esetting, Measurement, Settings, Eset, Ting

MultiLab 4010-2
Settings
ba76142e01
05/2013
73
9.6
Reset
You can reset (initialize) all sensor settings and sensor-independent settings
separately from each other.
9.6.1 Resetting the measurement settings
pH
The following settings for pH measurements are reset to the default settings
with the Reset function:
The sensor settings are reset under the Reset menu item in the menu for cali-
bration and measurement settings. To open the settings, display the required
measured parameter in the measured value display and press the <ENTER>
key.
ORP
The following settings for ORP measurements are reset to the default settings
with the Reset function:
The sensor settings are reset under the Reset menu item in the menu for cali-
bration and measurement settings. To open the settings, display the required
measured parameter in the measured value display and press the <ENTER>
key.
The calibration data are reset to the default settings together with
the measuring parameters. Recalibrate after performing a reset.
Setting
Default settings
Buffer YSI
Calibration interval
7 d
Unit for slope
mV/pH
Measured parameter
pH
pH resolution
0.001
mV resolution
0.1
Asymmetry
0 mV
Slope
-59.2 mV
Man. temperature
25 °C
Single-point calibration
off
Setting
Default settings
mV resolution
0.1
Man. temperature
25 °C