Xylem MultiLab 4010-2 User Manual

Page 74

Advertising
background image

Settings

MultiLab 4010-2

74

ba76142e01

05/2013

Dissolved oxygen

The following settings for D.O. measurements are reset to the default settings
with the Reset function:

The sensor settings are reset under the Reset menu item in the menu for cali-
bration and measurement settings. To open the settings, display the required
measured parameter in the measured value display and press the <ENTER>
key.

Conductivity

The following settings for conductivity measurements are reset to the default
settings with the Reset function:

The sensor settings are reset under the Reset menu item in the menu for cali-
bration and measurement settings. To open the settings, display the required
measured parameter in the measured value display and press the <ENTER>
key.

Setting

Default settings

Calibration interval

180 d

Check interval

60 d

Measured parameter

D.O. concentration

Relative slope (S

Rel

)

1.00

Salinity (value)

0.0

Salinity (function)

Off

Setting

Default settings

Calibration interval

150 d

Measured parameter

ϰ

Cell constant (c)

Depending on the connected mea-
suring cell:
0.475 cm

-1

(calibrated)

0.475 cm

-1

(set)

0.100 cm

-1

Temperature compensation

nLF

Reference temperature

25 °C

Temperature coefficient (TC) of the
linear temperature compensation

2.000 %/K

Multiplier for TDS

1.00