Operation

Cond 3110

20

ba75792e02

11/2008

Note
If the error message E3 appears, refer to chapter 6 W

HAT

TO

DO

IF

...

Stability control

During calibration, the stability control is automatically activated.

Note
This method of automatically determining the cell constant by
calibration in the 0.01 mol/l KCL control standard solution can only be
used for measuring cells with cell constants in the range 0.450 ...
0.500 cm

-1

or 0.800 ... 0.880 cm

-1

.

3

Immerse the measuring cell in the control standard solution,
0.01 mol/l KCI.

4

Start the calibration with <ENTER>. The determination of the
cell constant with stability control starts. The

display

indicator flashes until there is a stable signal.
The cell constant determined is displayed. The meter
automatically stores the cell constant.

5

Switch to the measuring mode with <ENTER>.
The determined cell constant is used.

1/cm

8

0 3

8

L

A

C

Tref25

Cal

1/cm

4

0 3

8

C

Tref25

Cal