Technical data

Cond 3110

36

ba75792e02

11/2008

Note
The accuracy values specified here apply exclusively to the meter. The
accuracy of the measuring cell has also to be taken into account.

Measuring ranges,

resolution

Variable

Measuring range

Resolution

SAL

0.0 ... 70.0
according to the IOT
table

0.1

T [°C]

− 5.0 ... + 105.0

0.1

T [°F]

+ 23.0 ... + 221.0

0.1

Cell constants

Cell constant C

Values

Can be calibrated in
the ranges

0.450 ... 0.500 cm

-1

0.800 ... 0.880 cm

-1

Adjustable

0.800 ... 0.880 cm

-1

0.475 cm

-1

(fixed)

Reference temperature

Reference
temperature

Values

Adjustable

20 °C (Tref20)
25 °C (Tref25)

Accuracy (± 1 digit)

Variable

Accuracy

Temperature of the test
sample

/ Temperature compensation

Nonlinear (nLF)

± 0.5 %

0 °C ... + 35 °C

according to EN 27 888

± 0.5 %

+ 35 °C ... + 50 °C
enhanced nLF function

SAL / range

0.0 ... 70.0

± 0.1

+ 5 °C ... + 25 °C

± 0.2

+ 25 °C ... + 30 °C

T [°C] / temperature sensor

NTC 30

± 0.1

PT 1000

± 0.1