3 setting the cell constant manually, Setting the cell constant manually, La c – Xylem COND3110 HANDHELD CONDUCTIVITY METERS User Manual
Page 22: Es u, Lc l e
Advertising

Operation
Cond 3110
22
ba75792e02
11/2008
4.3.3
Setting the cell constant manually
Note
The cell constant to be set must either be taken from the operating
manual of the measuring cell or is printed on the measuring cell.
Range
0.800 ... 0.880 cm
-1
3
If necessary, press <CAL> repeatedly until CAL and the last
calibrated cell constant is displayed.
4
Confirm the selection with <ENTER>.
The displayed cell constant is used.
The meter switches to the measured value display.
1/cm
8
0 3
8
L
A
C
Tref25
Cal
1
Press <CAL> repeatedly until USE CELL is displayed.
2
Confirm the selection with <ENTER> or <CAL
_
>.
The cell constant that was set last is displayed.
3
If necessary, press <CAL>
repeatedly until a cell constant in
the range 0.800 ... 0.880 cm
-1
is displayed.
L
C L
E
E
S
U