6 what to do if, What to do if, Chapter 6 w – Xylem COND3110 HANDHELD CONDUCTIVITY METERS User Manual
Page 33: 6what to do if

Cond 3110
What to do if...
33
ba75792e02
11/2008
6
What to do if...
Error message
OFL, UFL
Error message,
E3
Sensor symbol flashes
Display,
LoBat
Meter does not react to
keystroke
You want to know which
software version is in
the meter
Cause
Remedy
– Measured value outside the
measuring range
– Use suitable measuring cell
Cause
Remedy
– Measuring cell contaminated
– Clean cell and replace it if
necessary
– Calibration solution not suitable
– Check calibration solutions
Cause
Remedy
– Cleaning interval expired
– Recalibrate the measuring
system
Cause
Remedy
– Batteries almost empty
– Replace the batteries (see
)
Cause
Remedy
– Operating condition undefined
or EMC load unallowed
– Processor reset:
Press the <ENTER> and
<On/Off>
key
simultaneously
Cause
Remedy
– E. g., a question by the service
department
– Switch on the meter.
During the display test,
display the software version
with <ENTER>.