6 what to do if, What to do if, Chapter 6 w – Xylem COND3110 HANDHELD CONDUCTIVITY METERS User Manual

Page 33: 6what to do if

Advertising
background image

Cond 3110

What to do if...

33

ba75792e02

11/2008

6

What to do if...

Error message

OFL, UFL

Error message,

E3

Sensor symbol flashes

Display,

LoBat

Meter does not react to

keystroke

You want to know which

software version is in

the meter

Cause

Remedy

– Measured value outside the

measuring range

– Use suitable measuring cell

Cause

Remedy

– Measuring cell contaminated

– Clean cell and replace it if

necessary

– Calibration solution not suitable

– Check calibration solutions

Cause

Remedy

– Cleaning interval expired

– Recalibrate the measuring

system

Cause

Remedy

– Batteries almost empty

– Replace the batteries (see

section 5.1 M

AINTENANCE

)

Cause

Remedy

– Operating condition undefined

or EMC load unallowed

– Processor reset:

Press the <ENTER> and
<On/Off>

key

simultaneously

Cause

Remedy

– E. g., a question by the service

department

– Switch on the meter.

During the display test,
display the software version
with <ENTER>.