2 using the last calibrated cell constant, Using the last calibrated cell constant, Es u – Xylem COND3110 HANDHELD CONDUCTIVITY METERS User Manual
Page 21: Lc l e

Cond 3110
Operation
21
ba75792e02
11/2008
Calibration evaluation
After the calibration, the meter automatically evaluates the current
status. The evaluation appears on the display.
Downloading calibration
data
You can download the calibration data.
4.3.2
Using the last calibrated cell constant
Precondition
A valid calibration must be available (see section 4.3.1).
Display
Cell constant [cm
-1
]
You are working with a
correctly calibrated measuring
cell.
in the range
0.450 ... 0.500 cm
-1
0.800 ... 0.880 cm
-1
E3
Eliminate the error according to
chapter 6 W
outside the ranges
0.450 ... 0.500 cm
-1
or
0.800 ... 0.880 cm
-1
1
Press <CAL
_
> to display the calibration data.
The calibrated cell constant is displayed.
1
Press <CAL> repeatedly until USE CELL is displayed.
2
Press <ENTER> or <CAL
_
> to confirm the selection of USE
CELL.
L
C L
E
E
S
U