Cond 3110

Operation

21

ba75792e02

11/2008

Calibration evaluation

After the calibration, the meter automatically evaluates the current
status. The evaluation appears on the display.

Downloading calibration

data

You can download the calibration data.

4.3.2

Using the last calibrated cell constant

Precondition

A valid calibration must be available (see section 4.3.1).

Display

Cell constant [cm

-1

]

You are working with a
correctly calibrated measuring
cell.

in the range
0.450 ... 0.500 cm

-1

0.800 ... 0.880 cm

-1

E3

Eliminate the error according to
chapter 6 W

HAT

TO

DO

IF

...

outside the ranges
0.450 ... 0.500 cm

-1

or
0.800 ... 0.880 cm

-1

1

Press <CAL

_

> to display the calibration data.

The calibrated cell constant is displayed.

1

Press <CAL> repeatedly until USE CELL is displayed.

2

Press <ENTER> or <CAL

_

> to confirm the selection of USE

CELL.

L

C L

E

E

S

U