7 technical data, 1 general data, 2 measuring ranges, resolution, accuracy – Xylem COND3110 HANDHELD CONDUCTIVITY METERS User Manual
Page 35: Technical data, General data, Measuring ranges, resolution, accuracy, Chapter 7 t, Echnical, Data, 7technical data

Cond 3110
Technical data
35
ba75792e02
11/2008
7
Technical data
7.1
General data
7.2
Measuring ranges, resolution, accuracy
Dimensions
approx. 240 x 190 x 80 mm
Weight
approx. 0.4 kg
Mechanical structure
Type of protection
IP 67
Electrical safety
Protective class
III
Test certificates
CE
Ambient
conditions
Storage
- 25 °C ... + 65 °C
Operation
-10 °C ... + 55 °C
Climatic class
2
Power
supply
Batteries
4 x 1.5 V alkali-manganese batteries,
type AA
Rechargeable
batteries
4 x 1,2 V NiMH rechargeable batteries,
type AA
(no charging function)
Operational life
Approx. 1000 h operating hours (batteries)
Guidelines
and norms used
EMC
EC directive 2004/108/EC
EN 61326-1
EN 61000-3-2
EN 61000-3-3
FCC Class A
meter safety
EC directive 2006/95/EC
EN 61010-1
Climatic class
VDI/VDE 3540
IP protection
EN 60529
Measuring ranges,
resolution
Variable
Measuring range
Resolution
[µS/cm]
0.0 ... 199.9
200 ... 1999
0.1
1
[mS/cm]
2.00 ... 19.99
20.0 ... 199.9
200 ... 1000
0.01
0.1
1