2 resetting all meter settings, Resetting all meter settings, Ti n i – Xylem COND3110 HANDHELD CONDUCTIVITY METERS User Manual
Page 30

Operation
Cond 3110
30
ba75792e02
11/2008
4.6.2
Resetting all meter settings
This function resets all meter settings to the default condition. The
relevant values are given in the following sections:
The following settings are also reset to the default condition:
Resetting the meter
settings
Note
The measuring system is possibly not calibrated after a reset. Before
measuring, make sure the meter uses the cell constant suitable for the
measuring cell.
System settings
Measurement settings
Setting
Default settings
Measured parameter
mS/cm or µS/cm
Adjusted cell constant
0.840 1/cm
1
Switch on the meter with <On/Off>.
The display test appears briefly on the display.
2
During the display test, press <M> to open the menu for the
reset of the meter settings.
Init is displayed.
3
Press <
▲><▼> to display no or YES.
YES: Reset the meter settings.
no: Retain the meter settings.
4
Confirm with <ENTER>.
The menu is finished.
The meter switches to the measuring mode.