6.2 Manual Operation Mode

In the Manual Operation Mode, once the voltage and current levels have been set, the unit wi11 keep these settings even after unit has been
switched off until further reset.
There are 3 ways to input a setting:
1. By the Jog Dial.
2. By the UP or DOWN key.
3. By the Key Pad.
For any wrong input, e.g. out of specification values, display will show "OUT OF RANGE" FOR ABOUT 3 SECONDS, then returns to
it's previous settings and ready for correct new input.

Manual Setting of Voltage and Current by Jog Dial

1. To change the voltage setting, just rotate the Jog Dial, no other step is needed.
2. To change the current setting, press ENTER twice, the LCD will show I-set, push for desired value.
3. Press ENTER to confirm your chosen current setting.

Manual Setting of Voltage and Current by Up & Down Key

1. To change the voltage setting, just press the Up & Down Key, no other step is needed.
2. To change the current setting, press ENTER twice, the LCD will show I-set, push the Up & Down key for desired value.
3. Press ENTER to confirm your chosen current setting.

Manual Setting of Voltage and Current by Key Pad

1. Press ENTER.

Display will show V -set and the present voltage setting, key in the desired voltage level.

2. Press ENTER to confirm your chosen voltage setting and the display will show I-set.
3. You can key in the desired current setting now, if you do not want to change the current setting, you still need to press ENTER again

to confirm the new settings of voltage and current to complete the operation.

4. You must press the ENTER key 3 times in total even if you want to change the voltage, otherwise the unit will flash CANCEL