1. IMPORTANT SAFETY INSTRUCTIONS


1.

Do not use this apparatus near water.

2.

Clean only with dry cloth.

3.

Do not block any ventilation openings.

4.

Do not install unit near any heat source or heating emitting devices.

5.

Prevent the power cord from being walked on or pinched.

6.

Unplug this unit during lightning storms or when unused for long periods of time.

PRECAUTIONS

1.

The unit must be used within its specified range.

The rated input voltages can be found on the rating label at the back of the unit.
Before plugging into the AC supply, check with the rating label.

2.

This unit has a built-in Tracking O.V.P. (Over Voltage Protection) feature. In the event of the output voltage becoming 10% greater than the set
value, the O.V.P. wi11 be triggered and the output power will be cut off and > FAULT<

warning will appear.

When you get this warning, switch off the unit and remove all loading.
If you switch the unit back on again it should resume normal operation. In the event that this problem persists contact the manufacturer.

3.

This unit has a buzzer built inside. The buzzer will sound when Over Temperature/Overload/Over Voltage Protection has been triggered.
When you get this warning, switch off the unit and remove all loading.
Check your load and output voltage setting.
Allow the unit to cool down for 30 minutes.
If you switch the unit back on again, it should resume normal operation.
In the event this problem persists, the unit must be investigated by your agent.

4.

Only use the supplied software and optional accessories with this unit.

5.

Refer all servicing to manufacturer.

Warning!
The maximum output voltage for Model 1698 is up to 60Vdc.
It may be hazardous to touch metal part of the terminals.