6.2 Capture Single Shot Signal

Digital Storage Oscilloscope can easily be used to capture a

single-shot or unrepeatable signal. Perform the following

steps to capture a single-shot signal.

 Connect the channel 1 probe to the unknown signal.
 Press the trigger MENU key to display the TRIGGER

menu.

 Press the Source softkey to select CH1.
 Press the Mode softkey to select the Auto trigger mode.
 Adjust the vertical and horizontal controls to observe the

signal roughly and find out the right Trigger Type and

Trigger mode.

 Press the Type softkey from the TRIGGER menu page

to select Pulse trigger type.

 Press More 1/2 sofkey to display the TRIGGER menu

page 2/2.

 Press Mode softkey to select Normal Trigger mode.
 Press More 2/2 sofkey to display the TRIGGER menu

page 1/2.

 Press Pulse Mode softkey to select

(positive less

than).

 Rotate the entry knob (

) to set up the pulse width.