Performance Verification Procedure

243

Time Scale Accuracy

Equipment needed: Fluke calibrator (preferred) or signal generator

1. Connect your calibrator’s output to CH1 of the oscilloscope

by 50ohm feed thru connector. If recommended calibrator

is not available, connect alternative equipment as follows:

2. On the oscilloscope, the time base is switched to the sweep

speed under test.

3. Set up a 10 MHz sine wave output from calibrator.

4. Press the Auto button on oscilloscope to get a stable

waveform.

5. Press the Measure button, soft key Time, and then soft key

Frequency.

6. Press the Acquire button and soft key Mode to set

Average mode.

7. Adjust average to 8. It should read 10 MHz.

8.

Switch sweep speed to 100 ms (or 200 ms) to have a

stable frequency reading. This reading should be less than

Oscilloscope

Calibrator or

Signal