4 alarm connections, Installation 3.8.4 alarm connections – Hach-Lange CL17 USER MANUAL User Manual

Page 24

Advertising
background image

22

Installation

3.8.4

Alarm connections

CAUTION
Fire hazard. Current to the relay contacts must be limited to 5 amps resistive. A
method to remove power from the relays locally must be available in case of an
emergency or for servicing the product. Power removal can be accomplished with
an external switch and a 5-amp fuse or with a switched 5-amp circuit breaker.

The analyzer contains two alarm relays designed for use with either high voltage (greater
than 30V-RMS and 42.2V-PEAK or 60 VDC) or low voltage (less than 30V-RMS and
42.2V-PEAK, or less than 60 VDC), but not a combination of both high and low voltage.
Refer to

3.8.5

,

Figure 11

, and

Figure 15

along with the instructions below for connection

information.

The Normally Open and Common relay contacts will be connected when an alarm
condition is active.

The relay connector accepts 18-12 AWG wire. Wire gauge should be determined by load
application. Wire gauge less than 18 AWG is not recommended.

1. Make sure no power is supplied to the instrument.

2. Strip the insulation on each wire back ¼-inch. Refer to

Figure 14

.

3. Remove the connector from the instrument if desired.

4. Insert the wire ends into the connector until the insulation seats against

the connector. Refer to

Figure 14

. (Do not seat the insulation under the terminal pinch

plates.)

5. Replace the connector if necessary and supply power to the instrument.

Figure 13 Voltage selector switch and fuse replacement

1

Voltage selection switch (SW1). Factory set at 115V.

4

Hot / Black Ø1

2

GND

5

Fuses (F1, F2) (T, 2.5A, 250V)

3

Neutral / White Ш2