Section 1

Specifications

Specifications are subject to change without notice.

General

Display

LCD, 3½-digit measurement readout and six-character alphanumeric scrolling
text line.

Enclosure

IP62-rated with the gasketed door latched

Instrument description

31.4 cm (W) x 45.4 cm (H) x 17.9 (D) (12.4 x 17.88 x 7.06 in.)

Mounting

Wall mount

Instrument shipping weight

10.49 kg (23 lb)

Warranty

Hach Company warrants its products to the original purchaser against any
defects that are due to faulty material or workmanship for a period of one year
from the date of shipment unless otherwise noted in the product manual.

Certification

CE approved. Listed to UL and CSA safety standards by ETL.

Sample Requirements

Sample flow rate to sample conditioning

200 to 500 mL/min

Inlet pressure to instrument

1 to 5 psig; 1.5 psig is optimum. Exceeding 5 psig can cause sample
tubing failure unless sample conditioning is used.

Inlet pressure to sampling conditioning

1.5 to 75 psig (with sample tube level with the bottom of the instrument—see

Figure 7 on page 16

.)

Sample temperature range

5 to 40 °C (41 to 104 °F)

Inlet fitting at instrument

¼-inch OD polyethylene tubing with quick-disconnect fitting

Drain fitting

½-inch hose barb

Sample conditioning

Use provided sample conditioning

Regent/Standard Requirements

Maximum reagent usage

One-half liter per month (each of the two reagents)

Reagent containers

High-density polyethylene (2) ½-liter bottles

Reagent containment

Reagent bottles are contained inside the analyzer enclosure and are vented
externally.

Electrical

Power requirements

100–115/230 VAC (selection switch inside instrument); 90 VA, 50/60 Hz, 2.5 Amp
fuse

Power connection

Connection made by three wire barrier terminal block through a ½-inch conduit
hole in the case. Wire range: 18 AWG.

Installation category

II

Polution degree

2

Protection class

I

Alarm relay outputs

Two unpowered SPDT relays each rated at 5 A resistive, 240 V ac maximum.
Can function as sample set point alarms (high or low) or as a system warning
indicator or a system alarm indicator. Use either high voltage (greater than 30 V
RMS and 42.2 V PEAK or 60 V dc) or low voltage (less than 30 V RMS and 42.2
V PEAK or 60 V dc). Do not use a combination of high and low voltage.

Alarm connection

Connection made by a removable three wire plug through a ½-inch conduit hole
in the case. Wire range: 12–18 AWG.

Recorder output

One isolated recorder output, 4–20 mA (can be adjusted to 0–20 mA).
Recommended load impedance 3.6 to 500 ohms

Recorder output connections

Connection made by a removable three wire plug through a ½-inch conduit hole
in the case. Wire range: 12–22 AWG.