Operation

5.2.2 Alarm setup

This menu accesses two unpowered alarm relays (

AL1

and

AL2

),

RECALL WARNINGS

,

and

CLEAR WARNINGS

. The alarm relays may be assigned to one of the alarm functions

listed below. Only one function can be assigned to each relay.

LO—Alarm is triggered if concentration is less than or equal to the trip point. (0.0 mg/L
to 5.0 mg/L limits)

HI— Alarm is triggered if concentration is greater than or equal to the trip point. (0.0
mg/L to 5.0 mg/L limits)

SYSTEM ALARM (SA)—Activates an alarm when a system alarm occurs.

SYSTEM WARNING (SW)—Activates an alarm when a system warning occurs.

In addition, alarm status can be obtained and alarms can be cleared using this menu.

The relay is activated when an alarm condition occurs. Either relay may be activated on a
high chlorine alarm, a low chlorine alarm or when a System Warning or System Alarm
occurs. Alarm relays can be used to control chemical feeds as an on/off control, by using
them as a high or low alarm set point.

The two alarm relays can also be used to bracket a specific control range by adding a
double-pole double-throw auxiliary relay and wiring it to electrically latch. This allows the
pump to be turned on at a low level and off at a higher level—with the concentration
dropping down in between.

RELAY TEST

Press

ENTER

to momentarily turn the relays off, then sequentially turn on ALARM1 and ALARM 2.

Automatically clears both alarms to complete the test.

REF

1

Momentarily displays the last Reference A/D counts.

SAMPLE

1

Momentarily displays the last Sample A/D counts.

ZERO

1

Momentarily displays the last Zero A/D counts.

CYCLE TIME

1

Momentarily displays the Motor On time (in seconds).

POWER FAIL
WARNING

Sets the power fail warning to

ON

or

OFF

. Press

ENTER

then use the

ARROW

keys to toggle

between

ON

and

OFF

. Press

ENTER

to select. Press

EXIT

to accept changes and return to

SETUP

.

If the power fail warning is turned on and the power is interrupted, a system warning will be
generated when power is restored. The factory default is off.

DIAG OUTPUT

Pressing

ENTER

will momentarily indicate

ON

at the display and any optional LonWorks®

interfaces present will receive diagnostic data.

KEYBOARD
LOCKOUT

Press

ENTER

then use the

ARROW

keys to toggle between

ON

and

OFF

. Press

ENTER

to select the

displayed option. Press

EXIT

to return to the

SETUP

menu. If the keyboard is locked out, no editing

functions are allowed and all diagnostics are disabled except this one. The factory default is

OFF

.

LANGUAGE

English language is the factory default. Spanish, French, and German can be enabled.
Press

ENTER

then use the

ARROW

keys to move between the available options. Press the

ENTER

key when the language of choice is displayed. Press the

EXIT

key to accept changes and return to

SETUP

.

DEFAULT SETUP

Cold Start—Press

ENTER

to reset the instrument to factory defaults. All user-specified settings will

be lost. The display will show the software version number (V 1.2 for example) until the next
analysis is complete.

1

These functions are used by service technicians to troubleshoot the instrument.