Troubleshooting

7.3 System warnings

Recall active warnings using the

RECALL WARNINGS

function in the alarm menu. To clear

the warnings use the

CLEAR WARNINGS

function in the alarm menu.

Table 11 System warnings

Warning

What the Alarm Indicates

Alarm Display and

Instrument Indication

Resolution

POWER

Power failed. In default condition, this warning
is disabled.

POWER

Inspect wiring and circuit breakers
that deliver power to the CL17.

Low Signal

Reference measurement is less than 2000
A/D counts.

LOWSIG

Clean sample cell
Replace sample cell

Marginal

Gain

A calibration resulted in a gain correction
factor greater than 1.1 or less than 0.9. The
calibration standard and the value that the
operator is attempting to input do not match.
Re-test the calibration standard and perform a
new calibration on the CL17.

MARG G

Set CAL Defaults.
Verify standard value.
Recalibrate the instrument.

Marginal

Zero

A calibration resulted in an offset correction
factor greater than 0.2 and less than 0.5.

MARG Z

Set CAL Defaults.
Verify zero value.
Recalibrate the instrument.

Marginal

Offset

A high offset count was measured with the
LED off. Usually caused by light leaks.

MARG 0

Verify the colorimeter end caps
are firmly screwed in place.
Check for light leaks.