2 clean the instrument enclosure, 3 clean the colorimeter, Maintenance 6.2.2 clean the instrument enclosure – Hach-Lange CL17 USER MANUAL User Manual

Page 44

Advertising
background image

42

Maintenance

6.2.2 Clean the instrument enclosure

With the enclosure securely latched, use a soft cloth and a mild detergent to wipe the
outside of the enclosure. Do not allow moisture to enter the enclosure.

6.2.3 Clean the colorimeter

The colorimeter measuring cell may collect sediment or develop a film growth on the
inside walls. Cleaning with an acid solution and a cotton swab on a monthly basis is
recommended. Depending on sample conditions, it may be necessary to clean the cell
more often.

CAUTION
Chemical exposure hazard. The chemicals used in this procedure may be
hazardous if inappropriately handled or accidentally misused. Please read all
warnings on the reagent labels. Protective eye wear is always recommended when
contact with chemicals is possible.

Colorimeter cell cleaning procedure:

1. Press the

MENU

key, then press the down

ARROW

key until the

MAINT

menu

appears. Press

ENTER

.

2. Press the down

ARROW

key to scroll to the

CLEAN

option. Press

ENTER

.

3. Refer to

Figure 22

and locate the colorimeter.

4. Remove the rubber plug on the top of the colorimeter assembly.

5. Remove the magnetic stir bar using a bent steel paper clip as a retrieval tool. Refer

to

Figure 22

.

6. Wait until

CLEAN

flashes on the status line of the display, then fill the colorimeter with

19.2 N Sulfuric Acid Standard Solution (203832).

Note: Lower normality sulfuric acid will NOT be sufficient for cleaning the colorimeter.

7. Allow the sulfuric acid to stand in the colorimeter for 15 minutes.

8. Insert a wooden or paper cotton-tipped swab

*

into the opening and move it up and

down with a gentle scrubbing action, cleaning the interior surfaces of the colorimeter
cell. Clean the stir bar with the cotton swab (

Figure 23

).

9. Install the stir bar.

10. Replace the rubber plug on the top of the colorimeter assembly and make sure the

cover is on the viewing port of the colorimeter.

11. Press the

EXIT

key to immediately return to normal operation; the instrument will

return to normal operation automatically after 60 minutes.

*

Do not use a plastic swab when cleaning the colorimeter with sulfuric acid. The the acid will dissolve the plastic.