Maintenance

6.2

Unscheduled maintenance

6.2.1

Fuse replacement

The T, 2.5A, 250V fuse used in this instrument is used for both 115V and 230V operation.

DANGER
Electrocution hazard. Remove power from the instrument when removing or
installing a fuse.

DANGER
Fire hazard. For continued protection against fire, replace the fuse only with a fuse
of the same type and rating.

Fuse replacement procedure:

1. Make sure there is no power supplied to the instrument. the instrument power switch

does not remove power from the fuses, Power must be disconnected remotely to
remove power from the fuses.

2. Remove the customer access cover.

3. Locate the fuse holders (near the terminal strip in the customer wiring compartment).

Refer to

Figure 13 on page 22

.

4. Remove the two fuses (F1 and F2) and replace them with two new fuses with the

same specifications (T, 2.5A, 250V). Refer to

Parts and accessories on page 49

.

5. Reinstall the customer access cover and resupply power.

Table 7 Replacement Tubing Lengths for

Figure 21

Item

Description

Length (Qty)

From...

To...

Item Number

1

1/8"

ID, 1/4" OD

4.5 inches (1)

Pump Body Out

Colorimeter

43293-00

2

1/32" ID, 3/32" OD

7.0 inches (2)

Pump Body Out

Y-Fitting

44253-00

3

1/16" ID, 3/32" OD

2.0 inches (4)

Pump Body In

Pump Body Out

42717-00

4

0.062 ID, 0.125" OD

6.0 inches (2)

Reagent Bottle Cap

Pump Body In

42076-00

5

0.062" ID, 0.125" OD

6.0 inches (2)

Reagent Bottle Cap

Reagent Vent Fitting

42076-00

6

1/32" ID, 3/32" OD

7.0 inches (2)

Reagent Bottle Bottom

Reagent Bottle Cap

45524-00

7

1/8" ID, 1/4" OD

7.0 inches (1)

Sample Bypass Tee

Pump Body In

43293-00

8

1/8" ID, 1/4" OD

1.5 inches

Sample Bypass Tee

Sample Inlet Fitting

43293-00

9

1/4" OD x 0.04 W, Black

varies (1)

Sample Conditioning Out

Case Fitting

30616-00

10

1/2"

ID

varies (1)

Instrument Drain

Customer Drain

(not supplied)

11

1/32"

ID, 3/32"

OD

3 inches

Sample Bypass Tee

Drain fitting

44253-00

12

1/32" ID, 3/32" OD

1.0 inch (1)

Y-Fitting

Colorimeter

44253-00

13

0.500” ID, 11/16” OD

12 inches

Colorimeter

Drain Fitting

54108-00

Table 8 Drain, air purge, sample-in, and enclosure locations

A

Air purge, 0.01 CFM at 20 PSIG max, ¼ in. O.D. tube

B

Case drain, ½ in. I.D. tube

C

Sample drain, ½ in. I.D. tube

D

Sample in, 5 PSIG max, ¼ in. O.D. tube

E

Enclosure