32

Operation

5.2 Instrument menu structure

The major menus in the CL17 Analyzer consist of

ALARMS

,

RECRDR

,

MAINT

, and

SETUP

.

Press the

UP

or

DOWN

ARROW

keys to access the menus. The sections below give

information on the functions of each of the menus and the submenus within them.

5.2.1 Setup menu

Day-to-day analyzer functions are accessed from the

SETUP

menu. To enter the

SETUP

menu, press the

MENU

key, then use the

UP ARROW

key to scroll to the

SETUP

option.

Press

ENTER

.

Scroll through the

SETUP

menu options using the

UP

and

DOWN ARROW

keys. Press

ENTER

to select the displayed menu option. Arrow icons are displayed if the display is

editable. Press the

EXIT

key to back out of a menu and return to

SETUP

. The menu

information in

Table 6

is in the order it appears in the

SETUP

menu.

Table 6 Setup menu options

Menu option

Description

SIGAVG

The SIGAVG feature is used to average readings and prevent erratic recorder output signals.
Press

ENTER

, use the

UP

and

DOWN ARROW

keys to select 1, 2, 3, or 4, then press

ENTER

to

accept the displayed selection. Depending on the value selected, the last 1, 2, or 3 measurements
are averages and the reported concentration is the averaged values. Option 4 is a special filter,
which only reports the maximum value between the current reading and the last. This feature
reduces negative spikes due to bubbles and other abnormalities in the sample. The factory default
is 1. Press

EXIT

to return to the

SETUP

menu.

Note: A properly maintained instrument in good working condition will read within the accuracy
specifications presented in

Specifications on page 3

. The calibration function should never be used

to force an analyzer that is not working correctly to display a correct reading. Calibrations are
intended for very small adjustments only.

CAL ZERO

This feature is used to offset the instrument zero using a prepared zero standard. If the offset is
too large, a marginal zero (MARG Z) warning will occur. Refer to

Calibration on page 36

.

CAL STD

This feature is used to adjust the analyzer slope using a prepared standard. If the adjustment is
too large, the analyzer will not accept it and a marginal gain (MARG G) warning will occur. The
new standard will not be accepted and it will have to be checked again using another method to
establish a more exact reading. Refer to

Calibration on page 36

.

CAL DEFAULTS

(Calibration Default Settings) Press

ENTER

to return the calibration settings back to factory

defaults and set the Gain to 1.0 and the Offset to 0. Use this feature to remove calibrations that
are causing a MARG G or MARG Z to be displayed.

OFFSET

Press

ENTER

to momentarily display the current Zero correction factor. The factory default is 0.

GAIN

Press

ENTER

to momentarily display the current Gain correction factor. The factory default for gain

is 1, however, readings of 0.8 to 1.25 are acceptable.

DISPLAY TEST

Press

ENTER

to momentarily display all LCD segments to visually check that the complete display

is functioning.

RECMIN

Recorder signal test. Press

ENTER

to drive the Recorder output to its minimum (normally 4 mA)

and hold it there until the next analysis is complete.

RECMAX

Recorder signal test. Press

ENTER

to drive the Recorder output to its maximum (normally 20 mA)

and will hold it there until the next analysis is complete.
Note: To hold the signal longer, press the

MENU

key then the

DOWN ARROW

key to scroll to

MAINTENANCE. Press

ENTER

then scroll down to

HOLD OUTPUTS

and press

ENTER

again.

REC ADJUST

Press

ENTER

to modify the Recorder Minimum and Recorder Maximum values. Use this feature to

offset the 4 or 20 mA output signal to force a chart recorder to read zero or full scale in case of a
small discrepancy. Use the

RIGHT

and

LEFT ARROW

keys to select

RECMIN

or

RECMAX

then use

the

UP

and

DOWN ARROW

keys to adjust the value. Press

ENTER

to select. Press

EXIT

to accept

changes and return to

SETUP

.

Note: The numbers on the display are A/D counts and cannot be used to determine the recorder
setting.