Specifications

Optical

Light source

Class 1 LED (light emitting diode) with a peak wavelength of 520 nm; 50,000
hours estimated minimum life

Performance

Operating range

0–5 mg/L free or residual chlorine

Accuracy

±5% or ±0.03 ppm whichever is greater

Precision

±5% or ±0.01 ppm whichever is greater

Quantitation limit

0.09 ppm

Cycle time

2.5 minutes

Detection limit

0.03 ppm

Calibration

Uses default calibration curve

Power switch

User accessible power switch is required

Recorder

One 4-20 mA/0–20 mA

Alarm relay outputs

Two SPDT relays, 5A resistive load at 240 V ac. Can function as sample set
point alarms (high or low) or as a system warning indicator or a system alarm
indicator.

Optional external outputs

Hach AquaTrend

®

Network Interface

Environmental

Storage temperature range

-40 to 60 °C (-40 to 140 °F)

Operating temperature range

5 to 40 °C (41 to 104 °F)

Humidity

90% at 40 °C (90% at 104 °F)

Air purge (optional)

0.1 CFM instrument quality air at 20 psig maximum, ¼-inch OD tubing

Altitude

2000 m (6561 ft)