Specifications
Optical
Light source
Class 1 LED (light emitting diode) with a peak wavelength of 520 nm; 50,000
hours estimated minimum life
Performance
Operating range
0–5 mg/L free or residual chlorine
Accuracy
±5% or ±0.03 ppm whichever is greater
Precision
±5% or ±0.01 ppm whichever is greater
Quantitation limit
0.09 ppm
Cycle time
2.5 minutes
Detection limit
0.03 ppm
Calibration
Uses default calibration curve
Power switch
User accessible power switch is required
Recorder
One 4-20 mA/0–20 mA
Alarm relay outputs
Two SPDT relays, 5A resistive load at 240 V ac. Can function as sample set
point alarms (high or low) or as a system warning indicator or a system alarm
indicator.
Optional external outputs
Hach AquaTrend
®
Network Interface
Environmental
Storage temperature range
-40 to 60 °C (-40 to 140 °F)
Operating temperature range
5 to 40 °C (41 to 104 °F)
Humidity
90% at 40 °C (90% at 104 °F)
Air purge (optional)
0.1 CFM instrument quality air at 20 psig maximum, ¼-inch OD tubing
Altitude
2000 m (6561 ft)