A.2 adding the measurement to a channel, A.3 alarms and warnings, Network interface card – Hach-Lange CL17 USER MANUAL User Manual

Page 58

Advertising
background image

56

Network Interface Card

A.2 Adding the measurement to a channel

The measurement from the CL17 Analyzer must be associated with a channel on the
AquaTrend Interface so the measurement will be shown on the AquaTrend display.
Attach the measurement to a channel as follows:

3. Select

AQUATREND MENU

and press the

ENTER

key.

4. Select

DISPLAY SETUP

and press

ENTER

.

5. Select

MEASUREMENT TO CHANNEL

and press

ENTER

.

6. Select

ATTACH

and press

ENTER

.

7. Select a available channel on the AquaTrend and press

ENTER

.

8. Press

ENTER

again to select

ATTACH TO CL17.

9. Press the

MENU

key to return to the AquaTrend main display. The reading will be

displayed and updated every 2.5 seconds.

A.3 Alarms and warnings

If an alarm or warning condition exists at the CL17 Analyzer, the alarm or warning will be
indicated at the AquaTrend via the alarm annunciator. The actual alarm or warning must
be viewed and resolved at the CL17 Analyzer. The CL17 Analyzer alarms and recorder
outputs can also be connected to a Signal Output Module. Please refer to the Signal
Output Module Manual

.

A.4 Logging data using the serial input/output module

The CL17 Analyzer sends the chlorine reading (in ppm) to a SIO on the network, if one is
present. The service diagnostic data must be disabled before outputting data to the SIO:

1. Press the

MENU

key then the

DOWN ARROW

key until

SETUP

is displayed.

2. Press

ENTER

.

3. Press

ENTER

when

DIAG OUTPUT

is displayed.

4. Use the up or

DOWN ARROW

key to toggle the selection between

ON

and

OFF

then

press

ENTER

to select the displayed option. Press the

EXIT

key to return to the

measurement display.

The format of the data is illustrated below, where n is the SIO channel number for the
CL17. (The channel number (1–8) is assigned based on the order in which devices were
added to the network.) Please refer to the AquaTrend Interface Manual for information on
determining the channel number.

Note: In the following commands, n signifies the channel number.

[n]3.00,<CR><LF><NULL>

Additional diagnostic strings will be present if service diagnostics is enabled. These
diagnostics are for service personnel only.