BDA 387/2

25

C5

Mean Value calculation

In colour and colour difference measurements it is sometimes required to take the mean of

several single measurements. For mean value calculation, press key M . All following

measurements are stored for the mean value calculation.


• Press the M key to select mean value

calculation. Then you see the display

shown on the right.

• Place

spectro-color

®

on the sample.

• Trigger measurement with the measure

button.


• From the second measurement on, the

sigma key

∑ and Cl appear.


• After pressing the Cl key, the latest

measured value is deleted.

• Pressing

the

∑ key finishes the mean

value calculation and displays the result.

The standard deviation and coefficient of

variability are indicated in the print-out, too.

Mean value calculation is feasible with

reference and difference measurement.

Difference meas. 

 
 

 

 
Take mean value 
 
 
 
Continue: Measure 

 

× 

Difference meas. 

Σ

Cl 

 
Measure 2 
   
ΔE* =   0.41 
   
ΔL* =   0.39 
   
Δa* =  ‐0.04 
   
Δb* =  ‐0.12 
   Mi  =   0.03 