Hach-Lange SPECTRO COLOR d-8 User Manual

Page 28

Advertising
background image

BDA 387/2

29

4. Then

click

Start an

d Run... .

5. Now the following dialogue box is

opened. At „open“ type A:\TRANSCAL

and press

OK

. Then the installation

programm starts. Now choose the

language and the PC-interface for

communication with spectro-color

®

and

PC.

6.

Select item

File

in the menu bar and choose the command

Load calibration values from file .

7. Select the file with cali-

bration data (e.g

00001234.scd) and

confirm your choice with

OK

. Now the calibration

data appears in the left

part of the dialogue box.

8. Switch on the spectro-

color

®

.

9. In the menu bar select

Edit /

Remote ON

. Now

the display of spectro-

color

®

has to read Remote Mode.

10. In the menu bar select

Edit /

Send standard values to spectro-color to load the calibration

data into spectro-color

®

.

11. Finally select

Edit

/

Remote OFF

in the menu bar and press

End

.



Then calibrate the instrument as described in C1 Calibration.