Hach-Lange SPECTRO COLOR d-8 User Manual
Page 28

BDA 387/2
29
4. Then
click
Start an
d Run... .
5. Now the following dialogue box is
opened. At „open“ type A:\TRANSCAL
and press
OK
. Then the installation
programm starts. Now choose the
language and the PC-interface for
communication with spectro-color
®
and
PC.
6.
Select item
File
in the menu bar and choose the command
Load calibration values from file .
7. Select the file with cali-
bration data (e.g
00001234.scd) and
confirm your choice with
OK
. Now the calibration
data appears in the left
part of the dialogue box.
8. Switch on the spectro-
color
®
.
9. In the menu bar select
Edit /
Remote ON
. Now
the display of spectro-
color
®
has to read Remote Mode.
10. In the menu bar select
Edit /
Send standard values to spectro-color to load the calibration
data into spectro-color
®
.
11. Finally select
Edit
/
Remote OFF
in the menu bar and press
End
.
Then calibrate the instrument as described in C1 Calibration.