A product description, A1 concise operating instructions – Hach-Lange SPECTRO COLOR d-8 User Manual

Page 3

Advertising
background image

BDA 387/2

4

A Product

Description

A1

Concise Operating Instructions

To turn spectro-color

®

on and off, press the flush button on the top of the instrument. The

warning „Instrument not calibrated for 24

hours!“ appears after 24 hours without cali-

bration at the latest. Press the centre black key

yes to call the calibration function directly. If

you do not wish to calibrate, because e.g. the

calibration standard is not available, you can

also measure without previous calibration.

Calibration against standard

• Press key Ø in the main menu to switch

to the configuration menu.

• Select [Calibration] by pressing the ok

key.

 

• Measure with spectro-color

®

the black

calibration standard.

• Measure with spectro-color

®

the white

calibration standard.

Colour measurement:

• Press key C .
• The display reads: : C D65/10/SPIN .
• Place spectro-color

®

in perpendicular

position to the sample and trigger the

measurement with the measure key.

• After the measurement the display shows

the colour values in L*a*b*-system

Difference Measurement:

• Press key ΔC .
• Place spectro-color

®

in perpendicular

position to the reference sample and trigger

the measure by the measure key.

• Display of colour values; L*, a*, b*
• Place spectro-color

®

in perpendicular

position to the sample and trigger the

measure by the measure key.

• Display of colour difference values; default

setting: dE*, dL*, da*, db*

yes 

 

Instrument not 
calibrated 
for 24 hours! 
 
Calibrate now=Yes 
Continue: Measure 

 

× 

Standard no.xxxxx 

 
 

 

 
Calibration 
against black 
 
Continue: Measure 

 

× 

C  D65/10/SPIN 
L* a* b* (CIE) 

Grf 

DIM 

 

   L* = 83.03 
   a* =  0.09 
   b* =  7.82 

 

× 

 Reference 


MEM 

 

Measure button:  
Measure reference 
 
MEM : Reference 
taken from memory  

Ч 

Advertising