Hach-Lange AMTAX inter2 Basic User Manual User Manual

Page 110

Advertising
background image

English 110

SETTINGS (one-channel operation)

CONTRAST (Adjustable read-off angle of the display [temperature-dependent])
--> -90 to +90 °
LANGUAGE (Menu guide: DE, EN, ordered language)
STANDARD (Concentration of the standard solution used):
0.5 (AMTAX inter 2 2)
4.90–5.10 in 0.01 steps (AMTAX inter 2 20)
34.0–36.0 in 0.1 steps (AMTAX inter 2 80)
GAIN CORR
(Correction factor
--> 0.050–20.0 in adapted steps)
RANGE (mg/L)
Measurement range of the device (for information)
--> 0.02–2.00 / 0.1–20.0 / 2.0–80
CUR.RANGE (Measurement range for the current output
(20 mA = set value))
--> depending on device type 0.2–2.0 / 2.0–20.0 / 20–100 mg/L NH

4

-N

CURRENT (Current transfer 0–20 mA, 4–20 mA)
MALFUNCTION (current transfer in case of malfunction
--> 0 mA, 20 mA)
SEDITAX (Sample feed via this removal and
homogenization system
--> yes, no)
MEAS.INTERVAL (if [SEDITAX] = no)
(Period between two measurements 5 min, 10 min)
INTEGR. (Average value formation over a longer period with strongly fluctuating measurement values
--> off, 2–20 [measurement values])
MIN (Bottom limit value contact (related to current output 1)
--> 0.0–2000 mg/L in 0.1 steps
MAX (top limit value contact (related to current output 1)
--> 0.0–2000 mg/L in 0.1 steps)
LOCATION (Any designation
--> 9 characters, each A-Z, 0-9, /,.,-,,,+,*,(,),\Q,&,./.,$,#,",!)
DATE (Setting of the date)
TIME (Setting of the time)