Error messages

All notes and error messages show in the bottom left of the display.
Detailed error details can be found in normal cases in the [STATUS]
menu – there, the errors can also be acknowledged.

Order numbers

Further accessories

Error Definition

Measures

SIGNAL LOW

Cuvette not filled
Optics or cuvette
dirty
Lamp failure
Lamp voltage too low
Amplifier defective/not
adjusted
Concentration too high

If neither cleaning nor a lamp
change rectifies the error,
call customer service

SIGNAL HIGH

Cuvette not filled
Lamp voltage too high
Amplifier defective/not
adjusted

Check sample feed
Call customer service

TEST
SETTINGS

Invalid or not plausible
data in the device data
memory
Frequent occurrence
points to an empty
buffer battery

The device data must
be completely checked
(F1 key)
Call customer service

SET
TIME/DATE

Time and/or date have
invalid settings and
were reset to 1/1/93,
15:00

A correction is required

LEAK IN THE
SYSTEM

The moisture sensor
has detected fluid and
after 10 seconds
stopped all pumps

Rectify error, dry out system
and push F1 [MEAS]
(5 seconds) to start the system
again

EPROM ERROR
CALL SERVICE

Error in the
program memory

When this error appears again
after switch-off and another
switch-on, call customer
service

RAM ERROR
CALL SERVICE

Error in the
data storage

When this error appears again
after switch-off and another
switch-on, call customer
service

CALIBRATION
FACTOR

Standard solution or
reagents missing
or are not
OK

Check pumps, hoses and
solenoid valves
Check canister fill level

Item no.

Designation

DOC023.XX.03107

User Manual

Available on request

FILTRAX with 2 m unheated hose

Available on request

FILTRAX with 10 m unheated hose

Available on request

FILTRAX with 20 m unheated hose

Available on request

FILTRAX with 30 m unheated hose

LZX414.00.00000

Tank edge hardware for module carrier

LZX676.00

Control unit bracket

LZX408

VIEWTAX program for data analysis

Advertising