English, Measurement device technical data – Hach-Lange AMTAX inter2 Basic User Manual User Manual

Page 93

Advertising
background image

English 93

English

Measurement device technical data

Subject to changes.

Performance specifications

Measurement method

Indophenol blue method, derived from DIN 38406 E5

Measurement ranges

AMTAX inter 2 – 2:
AMTAX inter 2 – 20:
AMTAX inter 2 – 80:

0.02–2.00 mg/L NH

4

-N

0.1–20 mg/L NH

4

-N

2.0–80 mg/L NH

4

-N

Measurement uncertainty

AMTAX inter 2 – 2:
AMTAX inter 2 – 20/80:

±4 % of the measurement value ±0.02 mg/L NH

4

-N

±2 % of the measurement value ±0.02 mg/L NH

4

-N

Measuring interval

5 or 10 min (selectable with continuous sample feed)

Measured value display

Graphic display with curve diagram

Calibration

Automatic at selectable intervals

Inspection interval

6 months

Maintenance requirements Approximately 1 h / week, typical

Reagent stock

Reagents:
Null/standard solution:
Cleaning solution:

4–8 weeks depending on interval
12 months
3–6 months depending on interval

Outputs

Two current outputs:
Two limit-value contacts:
Interface:

0/4...20 mA, maximum 500 Ω
Potential-free, 24 V 1 A
Bus-capable (optional)

Ambient temperature

+5° C to +40° C

Power connector

230 V AC / 50 Hz

Power consumption

Approximately 150 VA

Dimensions: W x H x D

560 mm x 810 mm x 330 mm

Mass

Approximately 29 kg (without reagents)

Fuses

T 2.5 A L; 250 V (2 pieces)

Miscellaneous

Auto-cleaning, data logger, graphic display with curve diagram