Hach-Lange AMTAX inter2 Basic User Manual User Manual

Page 112

Advertising
background image

English 112

+CHANNEL 1 (Sample feed via this removal and homogenization sys-
tem
--> yes, no)

GAIN CORR
(Correction factor
--> 0.050–20.0 in adapted steps)
INTEGR. (Average value formation over a longer period with strongly
fluctuating measurement values
--> off, 2–20 [measurement values])
CUR.RANGE (Measurement range for the current output
(20 mA = set value))
--> depending on device type 0.2–2.0 / 2.0–20.0 / 20–100 mg/L NH4-N
CURRENT (Current transfer 0–20 mA, 4-20 mA)
MALFUNCTION (Current transfer in error case
--> 0 mA, 20 mA, off)
LOCATION (Any designation
--> 9 characters, each A–Z, 0–9, /,.,-,,,+,*,(,),\Q,&,./.,$,#,",!)

+CHANNEL 2 (Submenu for current output 1 (sample 1), menu items as per [+CHANNEL 1])
CHAN.1 / CHAN.2 (Number of measurements directly one behind another per channel
--> 0–10 measurements each for channel 1 and channel 2)
MEAS.INTERVAL (if [SEDITAX] = no)
(Period between two measurements 5 min, 10 min)
INTEGR. (Average value formation over a longer period with strongly fluctuating measurement values
--> off, 2–20 (measurement values))
MIN (Bottom limit value contact (related to current output 1)
--> 0.0–2000 mg/L in 0.1 steps
MAX (top limit value contact (related to current output 1)
--> 0.0–2000 mg/L in 0.1 steps)
DATE (Setting of the date)
TIME (Setting of the time)
PASSCODE (Passcode protection on activation
--> 4-digit with numbers 1–4 via the keys F1–F4)
CLEANING (Trigger automatic cleaning, no, 6 h, 12 h, 1D–7D, start time for each: hourly. Recommendation: 1D (once a day)
--> last Date of the last auto-clean)
CALIBRATE (Trigger automatic calibration
--> no 1D–7D,
start time for each: hourly. Recommendation: 3 days (every 3 days)
--> last Date of the last automatic calibration)
VERSION (Program version)

SETTINGS (2 channel operation)