E-pumping and, Section general maintenance, General maintenance – Hach-Lange AMTAX inter2 Basic User Manual User Manual

Page 119

Advertising
background image

English 119

During a cuvette change, perform the following work procedures with
care (!)!

Measurement ranges 20 mg / 80 mg

1.

Press one of the function keys F1–F4 for seconds and select the

[+SERVICE] menu.

2.

Pull off the hoses coupled with quick connectors.

3.

Take out the cuvette upward at centering pin.

4.

After the installation of a new cuvette, start a calibration procedure

with [CALIBRATE].

Measurement range 2 mg

1.

Press one of the function keys F1–F4 for three seconds and select

the [+SERVICE] menu.

2.

Pull off the hoses coupled with quick connectors from the cuvette.

3.

Loosen the knurled screws of the halogen lamp and remove

halogen lamp – Caution: hot!

4.

Unscrew the cuvettes/optics block and remove the cuvette cover.

5.

The cuvette can now be carefully removed together with the two

sponge rubbers.

6.

After the installation of a new cuvette, start a calibration procedure

with [CALIBRATE].

General maintenance

Cleaning of the device

Clean the device with a soft, damp cloth. For stubborn contamination,
you can use a standard cleaning agent.

Maintenance

The manufacturer recommends the conclusion of an inspection contract.
This contract extends the warranty periods to five years and guarantees
the takeover of all inspections and repair work by qualified specialist
personnel. Furthermore, experiences and improvements flow constantly
into the devices and can be used by the user.

Maintenance schedule

A maintenance schedule belongs to the scope of delivery of each
device. To observe the maintenance tasks and inspections
unconditionally necessary for complex analysis machines, in this
schedule, the following reagent change / maintenance tasks /
inspections are acknowledged and the next respective due date entered.

REPAIR SCHEDULE AMTAX inter 2

Number: Correction

Factor:

Commissioning on:

by:

Inspection and repair within the
inspection contract

Maintenance work by the user

Every 12 months
Replacement
Connection hoses

Null solution

Standard solution
for measurement
range
20 mg / 80 mg

Pump
cassettes

Every 6 months
Replacement
Sample dosing
hose

Reagent dosing
hoses
onward

Halogen lamp

Cleaning solution

Check
Signal level

Electronics

Cleaning
Optics

Every 3 months
Cleaning
Glass parts

Replacement
Valve hoses

Sample dosing
hose onward

Standard solution
for measurement
range 2 mg

Every 4–8 weeks
Replacement
Reagents A, B

Check
Stirrer vessel

Overflow vessel

Sample feed and
sample drain
hoses