Installation

Electrical Installation

13

Figure 9

Removing Voltage Barrier

3.3.1 Installation in Conduit

In hard-wired electrical applications, the power and safety ground service drops for the instrument
must be 18 to 12 AWG. See

Figure 10 on page 13

for strain relief and conduit opening sealing plug

information. See

section 3.3.3 on page 14

for wiring information.

3.3.2 Installation Using a Power Cord

A sealing-type strain relief to maintain the NEMA 4X/IP66 environmental rating and a power cord
less than 3 meters (10 feet) in length with three 18-gauge conductors (including a safety ground
wire) can be used, see

Replacement Parts on page 39

. See

Figure 10 on page 13

for strain relief

and conduit opening sealing plug assembly. See

section 3.3.3 on page 14

for wiring information.

Figure 10

Using the Optional Strain Relief and Conduit Plug

1.

High voltage barrier

2.

Unsnap the barrier latch then pull out to remove the barrier.

NC

NC

NC

COM

COM

COM

NO

F1

F2

NO

NO

RELAY C

RELAY B

RELAY A

RELAY A

1

1

+ DATA

+ DATA

+ OUT 2

+ OUT 2

– DATA

– OUT 2

SERVICE REQUEST

SHIELD/CHASSIS GND

+ V

+ V

+ OUT 1

+ OUT 1

GND

– OUT 1

2

2

3

3

4

4

5

5

6

PROBES

ANAL

OG OUTP

UTS

PCB

CONNECTOR

PCB

CONNECTOR

FIELD WIRING

INSULATION MUST

BE RATED TO

80° C MINIMUM

FIELD WIRING

INSULATION MUST

BE RATED TO

80° C MINIMUM

DANGER - EXPLOSION HAZARD

DANGER - RISQUE D'EXPLOSION

DO NOT DISCONNECT WHILE CIRCUIT IS LIVE

UNLESS AREA IS KNOWN TO BE NON-HAZARDOUS.

NE PAS DEBRANCHER TANT QUE LE EST SOUS

TENSION, A MONIS QU'IL NE S'AGISSE D'UN

EMPLACEMENT NON-DANGEROUX

J1

J2

J4

NETWORK

INTERFACE

CARD

J3

J5

J6

U5

U9

S1

NC

NC

NC

COM

COM

COM

NO

F1

F2

NO

NO

RELAY C

RELAY B

RELAY A

RELAY A

1

1

+ DATA

+ DATA

+ OUT 2

+ OUT 2

– DATA

– OUT 2

SERVICE REQUEST

SHIELD/CHASSIS GND

+ V

+ V

+ OUT 1

+ OUT 1

GND

– OUT 1

2

2

3

3

4

4

5

5

6

PROBES

ANAL

OG OUTP

UTS

PCB

CONNECTOR

PCB

CONNECTOR

FIELD WIRING

INSULATION MUST

BE RATED TO

80° C MINIMUM

FIELD WIRING

INSULATION MUST

BE RATED TO

80° C MINIMUM

DANGER - EXPLOSION HAZARD

DANGER - RISQUE D'EXPLOSION

DO NOT DISCONNECT WHILE CIRCUIT IS LIVE

UNLESS AREA IS KNOWN TO BE NON-HAZARDOUS.

NE PAS DEBRANCHER TANT QUE LE EST SOUS

TENSION, A MONIS QU'IL NE S'AGISSE D'UN

EMPLACEMENT NON-DANGEROUX

J1

J2

J4

NETWORK

INTERFACE

CARD

J3

J5

J6

U5

U9

S1

2

1

1.

Power cord strain relief

2.

Conduit strain relief

3.

Conduit opening sealing plug

2

3

Advertising