Modbus Protocol

51

Comm/Net Card

Baud

40171

Unsigned

Integer

1

R/W

Sets the Modbus baud rate (0 = 9600;

1 = 19200; 2 = 38400; 3 = 57600;

4 = 115200)

Comm/Net Card

Stop Bits

40172

Unsigned

Integer

1

R/W

Sets the number of stop bits (1,2)

Comm/Net Card

Data Order

40173

Unsigned

Integer

1

R/W

Sets the register data order for floats
(0 = Normal; 1 = Reversed)

Comm/Net Card

Min
Response
Time

40174

Unsigned

Integer

1

R/W

Sets the minimum response time
(0 to 30 sec)

Comm/Net Card

Max
Response
Time

40175

Unsigned

Integer

1

R/W

Sets the maximum response time
(100 to 1000 sec)

Comm/Net
Card/Addresses

sc100

40176

Unsigned

Integer

1

R/W

Sets the sc100 Modbus Address

Comm/Net
Card/Addresses

Sensor 1

40177

Unsigned

Integer

1

R/W

Sets the Sensor 1 Modbus Address

Comm/Net
Card/Addresses

Sensor 2

40178

Unsigned

Integer

1

R/W

Sets the Sensor 2 Modbus Address

Comm/Net Card/Stats

Good
Messages

40179

Unsigned

Integer

2

R/W

Number of good messages

Comm/Net Card/Stats

Bad
Messages

40181

Unsigned

Integer

2

R/W

Number of failed messages

Comm/Net Card/Stats

% Good Mesg

40183

Float

2

R/W

% of good messages

Comm/Service Port

Mode

40185

Unsigned

Integer

1

R/W

Sets the Modbus mode (0 = RTU;
1 = ASCII)

Comm/Service Port

Baud

40186

Unsigned

Integer

1

R/W

Sets the Modbus baud rate (0 = 9600;
1 = 19200; 2 = 38400; 3 = 57600;
4 = 115200)

Comm/Service Port

Stop Bits

40187

Unsigned

Integer

1

R/W

Sets the number of stop bits (1,2)

Comm/Service Port

Data Order

40188

Unsigned

Integer

1

R/W

Sets the register data order for floats
(0 = Normal; 1 = Reversed)

Comm/Service Port

Min
Response
Time

40189

Unsigned

Integer

1

R/W

Sets the minimum response time
(0 to 30 sec)

Comm/Service Port

Max
Response
Time

40190

Unsigned
Integer

1

R/W

Sets the maximum response time
(100 to 1000 sec)

Comm/Service
Port/Addresses

sc100

40191

Unsigned
Integer

1

R/W

Sets the sc100 Modbus Address

Comm/Service
Port/Addresses

Sensor 1

40192

Unsigned
Integer

1

R/W

Sets the Sensor 1 Modbus Address

Comm/Service
Port/Addresses

Sensor 2

40193

Unsigned
Integer

1

R/W

Sets the Sensor 2 Modbus Address

Comm/Service
Port/Stats

Good
Messages

40194

Unsigned
Integer

2

R/W

Number of good messages

Comm/Service
Port/Stats

Bad
Messages

40196

Unsigned
Integer

2

R/W

Number of failed messages

Comm/Service
Port/Stats

% Good Mesg 40198

Float

2

R/W

% of good messages

Table 9: sc100 Modbus Registers (continued)