Installation

Connecting/Wiring the sc Sensor

19

Figure 16

Analog Output Connections

3.5 Connecting/Wiring the sc Sensor

DANGER
Explosion hazard. Do not connect or disconnect electrical components or circuits to the
equipment unless power has been switched off or the area is known to be non-hazardous.

3.5.1 Connecting the sc Sensor in a Non-hazardous Location

3.5.1.1 Attaching a sc Sensor with a Quick-connect Fitting

The sensor cable is supplied with a keyed quick-connect fitting for easy attachment to the
controller, see

Figure 17

. Retain the connector cap to seal the connector opening in case the

sensor must be removed. Optional extension cables may be purchased to extend the sensor cable
length. If the total cable length exceeds 100 m (300 ft), a termination box must be installed. When
used with the termination box, the maximum cable length is 400 m (1300 ft). See the Replacement
Parts section in the sensor manual for part number information.

NC

NC

NC

COM

COM

COM

NO

F1

F2

NO

NO

RELAY C

RELAY B

RELAY A

J1

J2

J4

S1

NETWORK

INTERFACE

CARD

J3

J5

J6

U5

U9

1

1

+ DATA

+ DATA

+ OUT 2

+ OUT 2

– DATA

– OUT 2

SERVICE REQUEST

SHIELD/CHASSIS GND

+ V

+ V

+ OUT 1

+ OUT 1

GND

– OUT 1

2

2

3

3

4

4

5

5

6

PROBES

ANAL

OG OUTP

UTS

ANAL

OG OUTP

UTS

PCB

CONNECTOR

PCB

CONNECTOR

FIELD WIRING

INSULATION MUST

BE RATED TO

80° C MINIMUM

FIELD WIRING

INSULATION MUST

BE RATED TO

80° C MINIMUM

DANGER - EXPLOSION HAZARD

DANGER - RISQUE D'EXPLOSION

DO NOT DISCONNECT WHILE CIRCUIT IS LIVE

UNLESS AREA IS KNOWN TO BE NON-HAZARDOUS.

NE PAS DEBRANCHER TANT QUE LE EST SOUS

TENSION, A MONIS QU'IL NE S'AGISSE D'UN

EMPLACEMENT NON-DANGEROUX

J5

J5

NC

NC

NC

COM

COM

COM

NO

F1

F2

NO

NO

RELAY C

RELAY B

RELAY A

J1

J2

J4

S1

NETWORK

INTERFACE

CARD

J3

J5

J6

U5

U9

1

1

+ DATA

+ DATA

+ OUT 2

+ OUT 2

– DATA

– OUT 2

SERVICE REQUEST

SHIELD/CHASSIS GND

+ V

+ V

+ OUT 1

+ OUT 1

GND

– OUT 1

2

2

3

3

4

4

5

5

6

PROBES

ANAL

OG OUTP

UTS

ANAL

OG OUTP

UTS

PCB

CONNECTOR

PCB

CONNECTOR

FIELD WIRING

INSULATION MUST

BE RATED TO

80° C MINIMUM

FIELD WIRING

INSULATION MUST

BE RATED TO

80° C MINIMUM

DANGER - EXPLOSION HAZARD

DANGER - RISQUE D'EXPLOSION

DO NOT DISCONNECT WHILE CIRCUIT IS LIVE

UNLESS AREA IS KNOWN TO BE NON-HAZARDOUS.

NE PAS DEBRANCHER TANT QUE LE EST SOUS

TENSION, A MONIS QU'IL NE S'AGISSE D'UN

EMPLACEMENT NON-DANGEROUX

J5

J5

J6

1
2
3
4
5

Advertising