9 system setup menu (continued), Operation – Hach-Lange SC 100 User Manual

Page 38

Advertising
background image

System Setup Menu
34

Operation

Function set to FEEDER CONTROL

PHASE—“High” phase assigns the relay setpoint to respond to an increasing measured value; conversely, a “Low” phase
assigns the relay setpoint to respond to a decreasing measured value.

SET SETPOINT—Sets the value where the relay will turn on.

DEADBAND—Sets the range where the relay remains on after the measured value decreases below the setpoint value (high
phase relay) or increases above the setpoint value (low phase relay).

OVERFEED TIMER—Sets the time (0–999,9 minutes) to limit how long the relay can remain on.

OFF DELAY—Sets a time (0–999 seconds) to delay the relay from normally turning off.

ON DELAY—Sets a time (0–999 seconds) to delay the relay from normally turning on.

Function set to EVENT CONTROL

SET SETPOINT—Sets the value where the relay will turn on.

DEADBAND—Sets the range where the relay remains on after the measured value decreases below the setpoint value (high
phase relay) or increases above the setpoint value (low phase relay).

OnMax TIMER—Sets the time (0–999 minutes) to limit the time the relay can remain on.

OffMax TIMER—Sets a time (0–999 minutes) to delay the relay from normally turning off.

OnMin TIMER—Sets the time (0–999 minutes) to limit the time the relay can remain on.

OffMin TIMER—Sets the time (0–999 minutes) to limit the time the relay can remain off.

Function set to TIMER (RTC selected in SELECT SOURCE)

HOLD OUTPUTS—Set OUTMODE to select output hold operation and select channels that cause the outputs to be held.

INTERVAL—Set the off time for the relay.

DURATION—Set the on time for the relay.

OFF DELAY—Set the time for additional hold/output time after the relay has been turned off.

Function set to WARNING CONTROL

WARNING LEVEL—Set the warning level that will trigger a relay. Range: 0–32. For example: If warnings 1–9 are active on
the instrument, set the warning level to 0 to allow all warnings to trigger the relay; set the warning level to 5 to allow warnings
6 and above to trigger the relay. Set the warning level to 9 or greater to not trigger the relay on any warning.

Function set to PWM CONTROL

SET MODE—Auto or Manual

PHASE—Direct or Reverse

SET SETPOINT—Control Setpoint

DEAD ZONE—Zone around setpoint where output is off

PERIOD—3–60 second PMW period

MIN WIDTH—Minimum pulse width in 0,1 seconds

MAX WIDTH—Maximum pulse width in 0,1 seconds

PROP BAND—Proportional control band

INTEGRAL—Integral control setting (minutes)

Function set to FREQ CONTROL

SET MODE—Auto or Manual

PHASE—Direct or Reverse

SET SETPOINT—Control Setpoint

DEAD ZONE—Zone around setpoint where output is off

MIN WIDTH—0,001–200 pulses per minute

MAX WIDTH—0,001–200 pulses per minute

PROP BAND—Proportional control band (this is outside of the dead zone)

INTEGRAL—Integral control setting (minutes)

4.9 System Setup Menu (continued)