Alarms and Relays
18

Installation

Figure 15

Alarm and Relay Connections

3.4.2

Connecting the Analog Outputs

DANGER
Explosion hazard. Do not connect or disconnect electrical components or circuits to the
equipment unless power has been switched off or the area is known to be non-hazardous.

Two isolated analog outputs (1 and 2) are provided, see

Figure 16

. Each output can be set to 0–20

or 4–20 mA, and can be assigned to represent the measured parameter or secondary
measurement such as temperature. Make connections with twisted-pair shielded wire and connect
the shield at the controlled component end or at the control loop end.

Do not connect the shield

at both ends of the cable. Use of non-shielded cable may result in radio frequency emission or
susceptibility levels higher than allowed. Maximum loop resistance is 500 Ohm. Refer to the
sensor manual for output software setup.

Make wiring connections at the analyzer end as shown in

Figure 16

.

NC

NC

NC

COM

COM

COM

NO

F1

F2

NO

NO

RELAY C

RELAY B

RELAY A

J1

J2

J4

S1

NETWORK

INTERFACE

CARD

J3

J5

J6

U5

U9

J5

J5

NC

NC

NC

COM

COM

COM

NO

F1

F2

NO

NO

RELAY C

RELAY B

RELAY A

J1

J2

J4

S1

NETWORK

INTERFACE

CARD

J3

J5

J6

U5

U9

J5

J5

J5

J6

J7

3

2

1

RELAY A

RELAY B

RELAY C

Disconnect
Power

NC

COM

NO

Table 4 Output Connections at Terminal Block J6

Recorder Wires

Circuit Board Position

Output 2 +

1

Output 2 –

2

Shield

3

Output 1 +

4

Output 1 –

5