Hach-Lange SC 100 User Manual

Page 7

Advertising
background image

Specifications

3

Section 1

Specifications

Specifications are subject to change without notice.

Component Description

Microprocessor-controlled measuring unit with measured value display, temperature display,
and menu-driven system

Controller Operating
Temperature

–20 to 60 °C (–4 to 140 °F); 95 % relative humidity, non-condensing with sensor/network
card load <7 W; –20 to 40 °C (–4 to 104 °F) with sensor /network card load <25 W

Controller Storage Temperature –20 to 70 °C (–4 to 158 °F); 95 % relative humidity, non-condensing

Enclosure

NEMA 4X/IP66 metal enclosure with a corrosion-resistant finish

Power Requirements

AC Powered sc100 model: 100–230 V AC ±10 %, 50/60 Hz; Power 15 W with 7 W
sensor/network card load, 37 W with 25 W sensor/network card load
24 V DC powered sc100 model: 24 V DC –15 % / +20 %; Power 16 W with 7 W
sensor/network card load, 34 W with 25 W sensor/network card load

Pollution Degree/
Installation Category

II; II

Outputs

Two (Analog (4–20 mA)) outputs, maximum impedance 500 Ohm. Optional digital network
connection. IrDA digital connection.

Relays

Three SPDT, user-configured contacts rated 100–230 V AC, 5 Amp resistive maximum for the
ac powered sc100 and 24 V DC, 5 A resistive maximum for the dc powered sc100.

Controller Dimensions

½ DIN—144 Ч 144 Ч 150 mm (5.7 Ч 5.7 Ч 5.9 inches)

Controller Weight

1 ,6 kg (3.5 lb)

Certifications

CE approved (with all sensor types)
Listed for use in general locations to UL and CSA safety standards by ETL
(with all sensor types)