Output Options
30

Operation

4.5 Output Options

Refer to the

System Setup on page 27

for more information on the output options menu.

The analyzer provides two isolated analog outputs (Output 1 and Output 2).

4.5.2 Hold/Transfer Outputs

During normal measurement operation, the analog outputs can be held at the last measured value
or transferred to a preset value. To hold or transfer the output until released:

4.5.1 Navigating to the Output Options Menu

Step

Select

Menu Level/Instructions

Confirm

1

MAIN MENU

2

SYSTEM SETUP

3

OUTPUT SETUP

4

SELECT OUTPUT

5

Customize the options, refer to

System Setup on page 27

.

6

Main Menu or Main Measurement Screen when Output options are configured.

Step

Select

Menu Level/Instructions

Confirm

1

MAIN MENU

2

TEST/MAINT

3

HOLD OUTPUTS

4

SET OUTMODE

5

Select HOLD OUTPUTS or XFER OUTPUTS

6

SET CHANNELS

7

Select ALL or one of the connected sensors.

8

ACTIVATION

9

Select LAUNCH

10

Main Menu or Main Measurement Screen

reading will

flash