5 output options, Operation, 2 hold/transfer outputs – Hach-Lange SC 100 User Manual
Page 34: 1 navigating to the output options menu

Output Options
30
Operation
4.5 Output Options
Refer to the
System Setup on page 27
for more information on the output options menu.
The analyzer provides two isolated analog outputs (Output 1 and Output 2).
4.5.2 Hold/Transfer Outputs
During normal measurement operation, the analog outputs can be held at the last measured value
or transferred to a preset value. To hold or transfer the output until released:
4.5.1 Navigating to the Output Options Menu
Step
Select
Menu Level/Instructions
Confirm
1
MAIN MENU
—
2
SYSTEM SETUP
3
OUTPUT SETUP
4
SELECT OUTPUT
5
Customize the options, refer to
System Setup on page 27
.
6
Main Menu or Main Measurement Screen when Output options are configured.
—
Step
Select
Menu Level/Instructions
Confirm
1
MAIN MENU
—
2
TEST/MAINT
3
HOLD OUTPUTS
4
SET OUTMODE
5
Select HOLD OUTPUTS or XFER OUTPUTS
6
SET CHANNELS
7
Select ALL or one of the connected sensors.
—
8
ACTIVATION
9
Select LAUNCH
10
Main Menu or Main Measurement Screen
reading will
flash