Error list, Warning list – Hach-Lange POLYMETRON 9500 Ultrapure pH_ORP Module User Manual User Manual

Page 16

Advertising
background image

Table 3 Sensor DIAG/TEST menu (continued)

Option

Description

POLARIZATION

Contacting conductivity sensors only. Shows
information about the electrode polarization, the cable
capacitance and the time before the next
measurement.

IMPED STATUS

For pH sensors only—Enable or disable impedance
measurements.

SENSOR SIGNALS

Shows the current sensor signal information.

SENSOR DAYS

Shows the number of days that the sensor has been in
operation.

RESET SENSOR

Resets the number of days that the sensor has been in
operation and resets all calibration data to defaults.

Error list

Errors may occur for various reasons. An error icon consists of an exclamation point within a circle.
When an error occurs, the error icon and the measurement screen flash alternately in the main
display. All outputs are held when specified in the controller menu. To view errors, push the menu
key and select DIAGNOSTICS. Then select the device to view any problems associated with that
device.
A list of possible errors is shown in

Table 4

.

Table 4 Error list for pH and ORP sensors

Error

Description

Resolution

ADC FAILURE

The analog to digital conversion
failed

Power off and power on the
controller. Call technical support.

SAME BUFFER

The buffers for 2 point buffer
calibration have the same value

Make sure the two buffers used for
calibration are different.

SENSOR MISSING

The sensor is missing or
disconnected

Examine the wiring and connections
for the sensor and for the module.

TEMP MISSING

The temperature sensor is missing

Examine the wiring for the
temperature sensor. Make sure that
the correct temperature element is
selected.

GLASS IMP LOW

The glass bulb is broken or reached
end of life

Replace the sensor. Call technical
support.

Warning list

A warning icon consists of an exclamation point within a triangle. Warning icons appear on the right
of the main display below the measurement value. A warning does not affect the operation of menus,
relays and outputs. To view warnings, push the menu key and select DIAGNOSTICS. Then select
the device to view any problems associated with that device. The warning icon will no longer be
displayed once the problem has been corrected or acknowledged.
A list of possible warnings is shown in

Table 5

.

Table 5 Warning list for pH and ORP sensors

Warning

Description

Resolution

PH TOO HIGH

The measured pH is > 14

Calibrate or replace the sensor.

ORP TOO HIGH

The measured ORP value is >
2100 mV