Table 5 Warning list for pH and ORP sensors (continued)

Warning

Description

Resolution

PH TOO LOW

The measured pH is < 0

Calibrate or replace the sensor.

ORP TOO LOW

The measured ORP value is
< –2100 mV

OFFSET TOO HIGH

The offset is > 8 (pH) or 200 mV
(ORP)

Follow the maintenance procedures
for the sensor and then repeat the
calibration.

OFFSET TOO LOW

The offset is < 6 (pH) or –200 mV
(ORP)

SLOPE TOO HIGH

The slope is > 60 (pH)/1.3 (ORP)

Repeat the calibration with a fresh
buffer or sample.

SLOPE TOO LOW

The slope is < 54 (pH)/0.7 (ORP)

Clean the sensor, then repeat the
calibration.

TEMP TOO HIGH

The measured temperature is >
100 °C

Make sure that the sensor is
configured for the correct
temperature element.

TEMP TOO LOW

The measured temperature is <
0 °C

CAL OVERDUE

The Cal Reminder time has expired

Calibrate the sensor.

NOT CALIBRATED

The sensor has not been calibrated

Calibrate the sensor.

REPLACE SENSOR

The sensor has been in operation >
365 days

Replace the sensor.

CAL IN PROGRESS

A calibration was started but not
completed

Return to calibration.

OUTPUTS ON HOLD

During calibration, the outputs were
set to hold for a selected time.

The outputs will become active after
the selected time period.

FLASH FAILURE

External flash memory failed

Contact technical support.